RecallDepth

Discovery MR450 1.5T, Nuclear Magnetic Resonance Imaging System

Company
Ge Healthcare, Llc
Recall Initiated
June 4, 2021
Posted
July 12, 2021
Recall Number
Z-2064-2021
Quantity
40 units
Firm Location
Waukesha, WI

Reason for Recall

GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation on the system prior to running the initial 3-Plane Localizer Scan (See Figure 1 for reference to buttons.) Selecting and saving a patient orientation that does not match the patient's actual position may result in incorrectly annotated and/or flipped images.

Distribution

Worldwide distributions.

Root Cause

Software design

Action Taken

Deployment of the original customer letter to consignees in the U.S. began on June 4, 2021. Dissemination of the updated customer letter to consignees in the U.S. began on June 11, 2021. The letters discuss an issue on the affected products listed below where the Patient Orientation button may inadvertently be clicked when intending to click on the Save RX button. Users are directed to review and confirm that the actual patient orientation matches the prescribed patient orientation before initiating the scan. A correction is planned.

More recalls by Ge Healthcare, Llc

View all recalls by this company →