RecallDepth

DEXCOM G6 SENSOR 3 PACK Item Number: 5447552

Company
Cardinal Health Inc.
Recall Initiated
March 16, 2021
Recall Number
Z-1934-2021
Quantity
2357 kits
Firm Location
Dublin, OH

Reason for Recall

Due to shipping delays the devices were exposed to temperature excursions of 31.9F, intermittently over a 2 hour and 15 minute period of time which may cause inaccurate results.

Distribution

US Nationwide distribution in the states of FL, GA, SC.

Lot / Code Info

NDC/UPC: 8627005303 All lots numbers shipped from 2/23/201 to 3/10/2021

Root Cause

Environmental control

Action Taken

On 03/16/2021, Cardinal Health issued an Urgent Medical Device Removal notice to customer via letter notifying them a recall is being initiated as a precautionary measure as a result of exposure to temperatures outside of the products labeling requirements. Actions Required 1. INSPECT your inventory for the affected product. 2. SEGREGATE and QUARANTINE all affected product. 3. RETURN the enclosed acknowledgment form via fax to 1-817-868-5362 or email to rxrecalls@inmar.com, even if you do not have any product on hand. 4. CONTACT Inmar to arrange for return and credit of any affected product: 1-855-696-2074 (9:00 AM to 5:00 PM, Monday thru Friday)