RecallDepth

Colpassist Vaginal Positioning Device - Product Usage: intended for use in general gynecological surgery to assist in the positioning and manipulation of the vagina during the procedure. Material No. M0068318210

Company
Boston Scientific Corporation
Recall Initiated
April 15, 2021
Recall Number
Z-1490-2021
Quantity
4,549 WW
Firm Location
Marlborough, MA

Reason for Recall

Potential for pinholes (sterile barrier breach) on the edge of the pouch to potentially render the device non-sterile, and may result in a post-operative infection

Distribution

Worldwide distribution - US Nationwide distribution and the countries of Algeria, Brazil, Colombia, Estonia, Germany, Hong Kong, Australia, Canada, Costa Rica, Finland, Great Britain, Italy, Austria, Chile, Czech Republic, France, Greece, Jordan, Belgium, Lebanon, Mauritius, Mexico, Morocco, Namibia, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Puerto Rico, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan.

Lot / Code Info

All unexpired lots: C003699 C003715 C003735 C003750 C003775 C003804 C003813 C003842 C003853 C003887 C003905 C003936 C003950 C003960 C003966 C003986 C003997 C004028 C004041 C004078 C004091 C004107 C004117 C004128 C004136 C004160 C004186 C004206 C004221  GTIN 08714729848097

Root Cause

Packaging process control

Action Taken

Boston Scientific issued Urgent Medical Device Product Removal on 4/15/21 via FedX. Letter states reason for recall, health risk and action to take; Individually packaged Upsylon Y Mesh (UPN M0068318200) is not impacted by this removal. This removal affects only the UPNs and batches listed below and in Attachment 1. No other UPNs or batches are impacted by this removal. Please cease further distribution or use of any remaining product affected by this removal immediately. If you identify any product from the affected lots/batches within your inventory, please segregate the product immediately and return it to BSC in accordance with the enclosed instructions and indicate on your Reply Verification Tracking Form the quantity of units from each lot/batch that you will be returning. BSC is notifying regulatory authorities of this removal as required. Please read carefully through the removal instructions included with this notification. Your local Sales Representative can answer any questions that you may have regarding this notification.

More recalls by Boston Scientific Corporation

View all recalls by this company →