BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P, REF: 169-97 containing software version 3.9.1.9

This recall is currently active, issued August 2, 2024. It was issued by Carefusion 303, Inc..

Company
Carefusion 303, Inc.
Recall Initiated
June 24, 2024
Posted
August 2, 2024
Recall Number
Z-2501-2024
Quantity
3 units
Firm Location
San Diego, CA
Official Source
View on FDA website ↗

Reason for Recall

Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.

Distribution

US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV.

Lot / Code Info

UDI: 10885403512568/ Serial Number: 16113307 16111232 15949570

Root Cause

Under Investigation by firm

Action Taken

On July 11, 2024, CareFusion, subsidiary of Becton, Dickinson and Company (BD) issued a "Urgent Medical Device Correction" Notification, via mail. On June 24, 2024, BD issued Customer Release Notes via their firm's customer portal. BD ask consignees to take the following actions: Clinicians: 1. When restocking the cabinet, refer to the MedBank restock workflow screen to verify the correct bin location rather than the printed label until the software update is provided. Users should verify the contents of the bin location being restocked. 2. Please circulate this notice to all those who require awareness within your organization or to any organization whom you may have transferred BD Pyxis" MedBank" CUBIE" Replenishment Station devices to. 3. Please promptly complete and return the enclosed Customer Response Form within 15 days to acknowledge receipt of this notification and that you understand the information provided in this letter Distributor: 1. Please provide a copy of the attached customer letter to all facilities within your distribution network that have any product listed in Appendix 1. 2. Please promptly complete and return the enclosed Distributor Response Form within 15 days to acknowledge receipt of this notification and that you understand the information provided in this letter.