Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray system

This recall is currently active, issued May 7, 2024. It was issued by Nrt X-Ray A/S Birkegardsvej 16 Hasselager Denmark.

Company
Nrt X-Ray A/S Birkegardsvej 16 Hasselager Denmark
Recall Initiated
April 17, 2024
Posted
May 7, 2024
Recall Number
Z-1784-2024
Quantity
35
Official Source
View on FDA website ↗

Reason for Recall

X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting the patient or operator.

Distribution

US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, WV, LA, NY, PA, HI, NV, WI, MN, MI.

Lot / Code Info

REF/UDI-DI/Serial Numbers: 03400000/05713464000527/627002, 627004, 627007, 639001, 639002, 639003, 639004, 639007, 639010, 647001, 647004, 647005, 648001, 651007, 658002, 658003, 658006, 658010, 666008, 675001, 675004, 682006, 682010, 682011, 696003, 696004, 696006, 696007, 696012. 03400010/05713464000121/647009, 651002, 651006, 666009, 682003, 682023

Root Cause

Device Design

Action Taken

On 04/17/24, correction notices were emailed to consignees who were asked to do the following: 1) If you experience the described uncontrolled motion, stop using the device and contact your service provider to have your device repaired. 2) The correction notice should be shared with anyone who needs to be aware within your organization and forwarded to any organization where potentially affected devices have been transferred. 3) Complete and return the response form to either support@nrtxray.com or servicesupport@alpha-imaging.com A firmware update will be made to all motor controllers in the system. The update will be performed by the Service Provider. Customers with recall-related questions can call 1-800-331-7327.