Trevo Trak 21 Microcatheter, Catalog: 90338
- Company
- Stryker Neurovascular
- Recall Initiated
- April 23, 2024
- Posted
- May 28, 2024
- Recall Number
- Z-1906-2024
- Quantity
- 899
- Firm Location
- Fremont, CA
Reason for Recall
Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.
Distribution
Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA, SC, GU, WA, MO, OR, DC, OK, LA, MD, ID, NH, SD, NV, RI, ND, HI, MS, AR, IA, NM, VT, KS and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, SAR China, Hungary, India, Ireland, Israel, Italy, Japan, Korea (South), Lithuania, Luxembourg, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Republic of China, Thailand, Turkey, United Arab Emirates, United Kingdom.
Lot / Code Info
Catalog: 90338, UDI-DI: 07613327313956, Lot: 119461, 119473, 119598, 120070, 120201, 120970, 140512, 136912, 136913, 136914, 136915
Root Cause
Process design
Action Taken
On 4/23/24, recall notices were mailed to Supply Chain Management, Recall Coordinators, and Inventory Managers who were asked to do the following: 1) Cease further distribution or use of all affected devices. 2) Segregate the affected units in a secure location for return to the firm. 3) Circulate this Recall-Removal notice internally to all interested/affected parties. 4) Maintain awareness of this communication internally until all required actions have been completed within your facility. 5) Inform the firm if any of the subject devices have been distributed to other organizations. If yes, provide contact details so the firm can inform the recipients appropriately. 6) Inform the firm of any adverse events concerning the use of the subject devices. (US Customer Service: 1-855-91-NEURO (1-855-916-3876), Email: NVCustomerCare@stryker.com) 7) Complete and return the response form via email to nvfieldactions@stryker.com