Route 92 Medical, 070 Access System, REF 1004, Rxonly , Sterile EO. Used for introduction of interventional devices into the neurovasculature.

Recall Initiated
March 8, 2024
Posted
April 24, 2024
Recall Number
Z-1612-2024
Quantity
117 units
Firm Location
San Mateo, CA

Reason for Recall

Their is the potential for distal tip separation at the proximal marker band of the delivery catheters

Distribution

US: AL AZ CA DE FL GA IA IL IN KY LA MA MD MI MN MO NC ND NJ NY OH OR PA RI SD TN TX UT VA WV OUS: New Zealand

Lot / Code Info

UDI: 00853799007442/ Lot # 23061601 24010306, 23081802, 24011801, 23100601, 24020801, 23102601

Root Cause

Process control

Action Taken

On March 12, 2024 Route 92, issued a Urgent Medical Device recall notification to affected consignees via Mail and EMail. Route 92 asked consignees to take the following actions: Ensure all of your SUMMIT MAX inventory including the affected 070 Reperfusion Systems is in a secure, quarantined location. 2. Complete the Acknowledgement Form in Attachment 2 and return to Route 92 Medical by one of the following methods: Email-FieldActions@route92medical.com/Standard Mail - Attention Line: Tenzing 7 fiedl action, 155 Bovet Road Suite 100, San Mateo, CA 94402 Please return the Acknowledgement Form even if you no longer have the product subject to this recall. 4. Immediately return any affected products with the specific lot numbers listed in Attachment 1 to the address above using the original shipping container or request a new shipping container and pre-printed shipping label by email. 5. Keep this notice until all product subject to this recall has been returned. While processing your returns, please maintain a copy of this notice with the quarantined product subject to this recall and keep a copy for your records. 6. Separately, as part of SUMMIT MAX closure, you will be sending back your remaining SUMMIT MAX inventory if you haven t done so already.