Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24

Company
Fujifilm Healthcare Americas Corporation
Recall Initiated
April 11, 2024
Posted
May 22, 2024
Recall Number
Z-1884-2024
Quantity
418 units
Firm Location
Lexington, MA

Reason for Recall

AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable).

Distribution

US Nationwide distribution.

Lot / Code Info

UDI-DI: (01) 04580292771704 Noblus Ultrasound Imaging System AC Adapter Serial Number Before K17160001 (Excluding K17160001)

Root Cause

Device Design

Action Taken

FujiFilm issued Urgent Medical Device Recall Letter via email on 4/11/24. Letter states reason for recall, health risk and action to take: Please complete and return the Field Action Verification Form attached on page 3 of this communication. Responding with the information requested is essential for ensuring appropriate action is taken. Please accept the replacement of AC adapter that is implemented the countermeasure. Please contact our field action coordinator, Kotei Aoki, via email at HCUSRegulatoryAffairs@fujifilm.com, if you have any further questions regarding this field action.