RecallDepth

Medivance Neonatal ArcticGel Pads, REF: 31802, 3180202, 31802C

This recall is currently active, issued March 26, 2024. It was issued by Medivance Inc..

Company
Medivance Inc.
Recall Initiated
February 16, 2024
Posted
March 26, 2024
Recall Number
Z-1381-2024
Quantity
1,654
Firm Location
Louisville, CO
Official Source
View on FDA website ↗

Reason for Recall

Neonatal pads are experiencing reduced water flow, which may reduce or prevent water heating or cooling performance of the pads that are part of a temperature management system.

Distribution

Worldwide - US Nationwide distribution including in the states of IL, AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Belgium, Australia, Singapore, South Korea, Taiwan, Japan, China, Thailand.

Lot / Code Info

REF/UDI-DI/Lot(Expiration Date): 31802/00801741132131/NGGX5390(31Oct2024), NGGX1179(31Oct2024), NGGX1156(31Oct2024), NGHP2415(28Feb2025), NGHP2411(28Feb2025), NGHT1139(31May2025), NGHTY603(31May2025), NGHV4383(31Jul2025), NGHW1507(31Aug2025); 3180202/10801741132138/NGGZ2622(31Dec2024), NGGV1829(31Aug2024), NGGW3254(30Sep2024), NGHQ2940(31Mar2025); 31802C/00801741223457/NGGV1784(31Aug2024)

Root Cause

Process control

Action Taken

On 2/16/24, BD mailed or emailed recall notices to customers who were asked to do the following: 1) Immediately Discontinue Use 2) Immediately quarantine affected pads, record your inventory totals on the Customer Response Form, discard all devices within your facility's control per your facility's procedures. 3) Share this notice with any users of the product within your facilities or with any facility affected product was distributed to. 4) Complete and return the attached Customer Response Form to the BD contact noted on the form. If you require further assistance contact BD Customer Support at 1-844-823-5433. Say "Recall" when prompted M-F 8am -5pm CT.