CereLink ICP Extension Cable (Extension Cable), Model No. 826845 Used with Codman CereLink Intracranial Pressure (ICP) Monitor, Model No. 826820/826820P. Cable is also provided with monitor as part of the same SKU.

This recall is currently active, issued December 14, 2023. It was issued by Integra Lifesciences Corp..

Company
Integra Lifesciences Corp.
Recall Initiated
October 5, 2023
Posted
December 14, 2023
Recall Number
Z-0548-2024
Quantity
4941
Firm Location
Princeton, NJ
Official Source
View on FDA website ↗

Reason for Recall

Firm is expanding the existing recall RES 90457 for the Cerelink, initiated in June 2022, to include all CereLink ICP Extension Cables. The updated corrective action includes a design change to the extension cable to address the "out of range readings" issue.

Distribution

US distribution nationwide to the following states: AL AR AZ CA CO CT DC DE FL GA IA KY LA MA ME MN ND NE NH NJ NY OK PA RI TN TX UT VA WI WV OUS distribution to Australia/New Zealand, Canada, and Europe/Middle East/Africa.

Lot / Code Info

UDI-DI 10381780520665 All lots distributed

Root Cause

Device Design

Action Taken

On October 5, 2023, the firm issued an updated letter, intended to replace the notification sent on June 22, 2022 for the out-of-range readings issue (RES 90457). As an output of the root cause and failure investigation, the design of the extension cable is being modified to incorporate a solution to address the "out of range readings" issue. The firm instructed customers to return all existing CereLink ICP Extension Cables in their inventory. This includes all CereLink ICP Extension Cables included with the monitor AND all CereLink ICP Extension Cables provided separately. Customers should check their inventory and quarantine any affected product. Once the firm receives the customer's response form, the firm will provide instructions to return the product. Customers with questions about this recall should contact their Account Manager.