EXALT Model D Single-Use Duodenoscope: Outer box UPN# M00542420 & Outer box UPN# M0054242CE0.

This recall is currently active, issued November 21, 2023. It was issued by Boston Scientific Corporation.

Company
Boston Scientific Corporation
Recall Initiated
September 27, 2023
Posted
November 21, 2023
Recall Number
Z-0396-2024
Quantity
US: 586 devices & OUS: 81 devices
Firm Location
Marlborough, MA
Official Source
View on FDA website ↗

Reason for Recall

Poor image quality due to fluid ingress in the lens.

Distribution

Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA VT, WA, WI & WV. The countries of Australia, Belgium, Canada, Czech Republic, Denmark, France, Germany, Great Britain, Italy, New Zealand, Spain, Sweden, Switzerland.

Lot / Code Info

UDI-DI (GTIN): 08714729983514, Outer box UPN# M00542420, Lot / Batch # 31137538, 31537335, 31228699, 31590604, 31257049, 31635405, 31257050, 31683854, 31258726, 31802917, 31329914, 31810037, 31398618, 31991955, 31438355, 32006891, 31473069, 32146684. UDI-DI (GTIN): 08714729995746, Outer box UPN# M0054242CE0, Lot / Batch # 31199348, 31712358, 31574944.

Root Cause

Process control

Action Taken

Notification of Urgent Medical Device Product Removal was issued via FedEx on Oct 3, 2023. The notice advises customers to immediately discontinue use and segregate affected product, and to complete and return the Reply Verification and Tracking Form (RVTF) even if you do not have any affected product. Customers are to post the notice near any affected product and notify end users. If further distributed, notify any customers to which product has been further distributed to. Reach out to your local Boston Scientific representative with any questions. The notice is batch / lot specific. Boston Scientific has implemented corrective changes and will make replacement EXALT Model D scopes available as soon as possible.