KWIKSTIK, P. venetum, Catalog numbers: a) 0794P (2-pack), b) 0794K (6-pack)
- Company
- Microbiologics Inc
- Recall Initiated
- October 18, 2023
- Posted
- November 15, 2023
- Recall Number
- Z-0307-2024
- Quantity
- 13 units
- Firm Location
- Saint Cloud, MN
Reason for Recall
A complaint investigation showed that the bulk pellets (lot 794-142) were contaminated with Aspergillus flavus.
Distribution
Worldwide distribution - US Nationwide and the countries of Italy, France, Turkey, Peru, Mexico.
Lot / Code Info
a) UDI/DI 20845357019169, Lot Numbers: 794-142-2, 794-142-4 b) UDI/DI 30845357019173, Lot Numbers: 794-142-3
Root Cause
Under Investigation by firm
Action Taken
Microbiologics issued an Urgent Medical Device Recall notice to its consignees on 10/18/2023 via email. The notice explained the issue and requested that the labs review their procedures to understand who the issue affects its usage, use or discard depending on its procedures, and contact Microbiologics if a replacement kit is needed. For questions or concerns contact the Recall Support team at 1.320.229.7073 or recall@microbiologics.com.