RecallDepth

Disposable Spetzler-Malis Bipolar forceps, 18CM SM DISP BAYONET 0.5MM TIP, Catalog Number 6760-180-005

Company
Stryker Corporation
Recall Initiated
September 11, 2023
Posted
November 7, 2023
Recall Number
Z-0239-2024
Quantity
2,321 total units
Firm Location
Portage, MI

Reason for Recall

An error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.

Distribution

US Nationwide. Australia, Canada, Japan, Korea, Singapore.

Lot / Code Info

GTIN: 7613327300192; Lot Numbers: 200507002

Root Cause

Under Investigation by firm

Action Taken

An URGENT Medical Device Recall notification letter was sent to customers on 9/11/23. Actions to be taken 1. Immediately check your inventory to locate and quarantine any affected products found at your facility. 2. Sign and return the enclosed Business Response Form (BRF) by email to Instruments.recalls@stryker.com, even if affected product is no longer in inventory. 3. Upon receipt of the completed BRF, a shipping label will be provided to return any recalled product on-hand and Stryker will contact you to arrange for replacement of your product(s). 4. Maintain awareness of this communication internally and inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients accordingly. Please contact Jordan Anderson at (269) 800-1347 or Instruments.recalls@stryker.com with questions or concerns. For reporting any serious adverse events or product quality problems contact Stryker at 1-800-253-3210, ext. 5555 8:00 A.M. to 7:00 P.M. (Eastern Time), Monday Friday or by email at inst.stryker.cs@stryker.com.

More recalls by Stryker Corporation

View all recalls by this company →