VentStar Anesthesia WT (P)180, MP00374

Company
Draeger Medical, Inc.
Recall Initiated
April 17, 2023
Posted
May 17, 2023
Terminated
September 3, 2024
Recall Number
Z-1444-2023
Quantity
130 units
Firm Location
Telford, PA

Reason for Recall

Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.

Distribution

Domestic: US Nationwide Distribution.

Lot / Code Info

UDI-DI: 04048675421948; All lots.

Root Cause

Process control

Action Taken

The firm distributed recall notifications to consignees via FedEx starting on 4/17/2023. Consignees are instructed to immediately stop use of affected devices, inspect current stock, and quarantine any affected devices to prevent use. Alternative therapy for patients should be instituted as soon as possible and ensure all users of affected devices within consignee organizations are aware of the recall, with copies being forwarded if devices were further distributed. To replace or return unused products, consignees are to call Draeger Customer Success between 8:00 AM - 6:00 PM EST, Monday thru Friday, at 1-800-437-2437 (option 2, then option 1), or via email at US-Medical@draeger.com. A Return Materials Authorization and pre-paid shipping label will be provided upon request. Any additional questions regarding the recall communication may be directed to Mr. Michael Kelhart between the hours of 8:00 AM - 4:30 PM by phone at 267-664-1131 or by email at mike.kelhart@draeger.com.