Lucid TCD 2.0 System, Model: NA-LUCM1. Transcranial Ultrasound System

Company
Novasignal Corp.
Recall Initiated
December 16, 2022
Posted
February 1, 2023
Recall Number
Z-1049-2023
Quantity
6
Firm Location
Los Angeles, CA

Reason for Recall

Transcranial Doppler/ultrasound system tablet has a faulty hard drive, which will cause the system to freeze and reboot. Initially, the system will reboot to a normal operable state but eventually the system will be rendered inoperable.

Distribution

US Nationwide and Worldwide Distribution: AZ, TX, SC, PA, IL. OUS: CA, SE, NL

Lot / Code Info

UDI-DI/Serial Numbers: 00850681007221/ 21020003, 20080004, 20100004, 21080001, 20120001; 00850681007009/ 19090005

Root Cause

Nonconforming Material/Component

Action Taken

On 12/16/22, recall notices were emailed to customers. Customers were told they would be contacted by the recalling firm to facilitate the proper disposition of affected devices. If an unexpected reboot occurs, the recalling firm recommends pausing the use of affected devices until unaffected devices arrive. Complete and return the customer response form. Customers with questions can contact Customer Operations at anchal@novasignal.com