VADER pedicle system, torque wrench, Catalog Number 42-703

Company
Icotec Ag Industriestrasse 12 Altstatten Sg Switzerland
Recall Initiated
November 28, 2023
Posted
January 10, 2024
Recall Number
Z-0701-2024

Reason for Recall

icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighten the nut screws during implantation was not functioning correctly. Due to the malfunction, the nut screws may not have been tightened sufficiently.

Distribution

US: AZ. MI, OH, CA, CO, TX, NY, NC, DE, PA

Lot / Code Info

220803-001, 220803-002, 220803-003

Root Cause

Other

Action Taken

icotec ag issued an URGENT Medical Device Recall notice to its consignees beginning on 11/28/2023 via email. The notice explained the problem with the device, the risk to health, and informed the medical facility that the affected product had been removed prior to the notice. Decision on whether patient-level follow-up is required should be made by the physician. Replacements will be provided where applicable. The firm expanded this recall on 12/27/2023 (notices delivered by US mail) to include additional lots and one additional product. Contact for questions: Laura Husken, laura.huesken@icotec.ch, Tel.: +41 78 438 78 73