EarlyVue VS30 Vitals Monitor, utilizing E01 (Oridion Microstream sidestream etCO2) and R21 (Masimo Rainbow RRa) simultaneously, Product Code 863380

This recall is currently active, issued September 29, 2023. It was issued by Philips North America Llc.

Company
Philips North America Llc
Recall Initiated
August 31, 2023
Posted
September 29, 2023
Recall Number
Z-2654-2023
Quantity
13 total devices
Firm Location
Cambridge, MA
Official Source
View on FDA website ↗

Reason for Recall

A software issue prevents the monitor from alarming when the high/low acoustic respiration rate (RRa) limits are violated.

Distribution

Worldwide - US Nationwide distribution in the states of MA, MI. The country of Brazil.

Lot / Code Info

UDI-DI: 00884838091412; Serial Numbers: DE72401360 DE72401393 DE72400231 DE72400232 DE72400235 DE72400254 DE72400237 DE72400238 DE72400239 DE72409250 DE72409240 DE72409249

Root Cause

Software design

Action Taken

An URGENT Medical Device Correction notification letter dated 8/29/23 was sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients or users If you use EarlyVue VS30 devices with both options E01 (Oridion Microstream Sidestream etCO2) and option R21 (Masimo Rainbow RRa): 1) Observe the EarlyVue VS30 monitor display screen for out-of-limit acoustic respiration rates (RRa) during procedures when there is simultaneous use of E01 (Oridion Microstream sidestream etCO2) and R21 (Masimo Rainbow RRa), while Respiration Source Mode is set to Auto, and the CO2 hardware module setting is switched to Standby. This notice should be passed on to all those who need to be aware within your organization or to any organization where EarlyVue VS30 with both options E01 (Oridion Microstream Sidestream etCO2) and option R21 (Masimo Rainbow RRa) devices have been transferred. Actions planned by Philips to correct the problem A Philips representative will contact you to schedule a visit from a Philips Field Service Engineer who will perform software upgrade to your device. If you need any further information, please contact your local Philips representative, or call 1-800-722-9377.