Medtronic ILLUMISITE Platform Console, Model Number ILS-1000-CS

Company
Covidien, Llc
Recall Initiated
May 22, 2023
Posted
June 26, 2023
Recall Number
Z-2008-2023
Quantity
1 unit
Firm Location
Minneapolis, MN

Reason for Recall

Single ILLUMISITE Platform Console was converted from a 220-240VAC model to a 120VAC configuration using a non-supplied power cord and non-supplied fuses in the field by Medtronic Field Service Engineer.

Distribution

International distribution in the country of Taiwan.

Lot / Code Info

UDI-DI: 10884521782211; Serial Number: 518503

Root Cause

Process control

Action Taken

Covidien notified the single consignee via letter on 05/18/2023. The consignee was instructed to quarantine and return affected product, forward the notification to those who need to be aware within the organization and to any location which the device has been transferred and complete and return the Customer Confirmation Form.