Giraffe OmniBed Carestation CS1, Model Numbers: a. 2082844-001-XXX, b. 2074920-001-XXXXXX, c. M1050205, d. M1135822, e. M1145475, f. M1145476, g. M1235957, h. M1149051, i. M1146098; Labeled as the following: GIRAFFE OMNIBED CARESTATION CS1, GIRAFFE OMNIBED CARESTATION, GIRAFFE OMNIBED, GIRAFFE OMNIBED CS1, OMNI G 115/230V SERVO2 CARESTATION, ASSY-LAU, OMNIBED CHINA EBASE S HUMID, Manufacturing assembly - Make, OBSOLETE-Omnibed Czech Servo O2 Nist/Scale/Re-usable Probe/SW w/port hole/CE Plug, ASSY-LAU, Omnibed Czech S-PH w/Nist O2 w/Reus Probe CE Power Cord, Finished good - Make, ASSY-LAU, Omnibed Czech S-PH w/Reusable Probe CE Power Cord, Finished good - Make, Omnibed Czech Servo O2 Nist/Scale/Re-usable Probe/SW w/port hole/Ce Plug, ASSY-LAU, Omnibed Indonesian S-GR W/Reuseable Probe CE Plug, Finished good - Make, ASSY-LAU, Omnibed ENG S-Gr w/REUProbe CE Power Cord, Finished good - Make, ASSY-LAU, Omnibed Portuguese S-PH w/NIST O2w/Reusable Probe CE Power Cord, Finished good - Make. Incubator, Neonatal, Warmer, Infant radiant

This recall is currently active, issued July 6, 2023. It was issued by Datex--Ohmeda, Inc..

Company
Datex--Ohmeda, Inc.
Recall Initiated
May 19, 2023
Posted
July 6, 2023
Recall Number
Z-2100-2023
Quantity
4,540 units
Firm Location
Wauwatosa, WI
Official Source
View on FDA website ↗

Reason for Recall

GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices

Distribution

Worldwide distribution.

Lot / Code Info

a. Model Number 2082844-001-XXX; GTIN Numbers: 010084068211686221, 010084068211686221; All Lot/Serial Numbers. b. Model Number 2074920-001-XXXXXX; No GTIN Number; Serial Numbers: HDGR52137, HDGL50099, HDGT60164, HDGT60143, HDGT60142, HDGT60139, HDGT60138, HDGT60137, HDGU60753, HDGT60652, HDGU60069, HDGT60141, HDGU60156, HDGU60120, HDGU60119, HDGT60163, HDGT60703, HDGT60245, HDGT60160, HDGT60459, HDGT60246, HDGT60235, HDGU67645, HDGU60065, HDGU60063, HDGU60061, HDGU63122, HDGU63121, HDGU60062, HDGU57534, HDGU67533, HDGU67534, HDGT60460, HDGU67236, HDGT60249, HDGU60201, HDGU60570, HDGU60620, HDGU60023, HDGU60020, HDGT60612, HDGU60022, HDGU60571, HDGU60686, HDGU60202, HDGV60190, HDGW60244, HDGU67677, HDGU67676, HDGU67675, HDGU67280, HDGU67416, HDGU67415, HDGU67382, HDGU67377, HDGU67369, HDGU67366, HDGU67362, HDGU67289, HDGU67284, HDGU67282, HDGU63165, HDGU63167, HDGU63166, HDGV61084, HDGU67436, HDGU67423, HDGU67422, HDGU67421, HDGU67420, HDGU67419, HDGU67418, HDGU67414, HDGU67413, HDGU67381, HDGU67380, HDGU67379, HDGU67378, HDGU67370, HDGU67368, HDGU67367, HDGU67365, HDGU67363, HDGU67361, HDGU67360, HDGU67359, HDGU67358, HDGU67357, HDGU67304, HDGU67303, HDGU67302, HDGU67301, HDGU67300, HDGU67299, HDGU67298, HDGU67297, HDGU67296, HDGU67295, HDGU67294, HDGU67293, HDGU67292, HDGU67291, HDGU67290, HDGU67288, HDGU67287, HDGU67286, HDGU67285, HDGU67283, HDGU67281, HDGJ51159, HDGT60613, HDGT60614, HDGU60700, HDGU60703. c. Model Number M1050205; No GTIN Number; Serial Numbers: HDGV60148, HDGU60656, HDGU60658, HDGT60481, HDGS50047, HDGV60146. d. Model Number M1135822; No GTIN Number; Serial Numbers: HDGL51737, HDGL51738, HDGL51734, HDGM50736, HDGM50841, HDGL51735. e. Model Number M1145475; No GTIN Number; Serial Numbers: HDGM51813, HDGM52120, HDGM52119, HDGN50474, HDGN50472, HDGN50471, HDGN50469, HDGN50466, HDGN50464, HDGN50848, HDGN50847, HDGN50846, HDGN50845, HDGN50844, HDGQ50049, HDGR53203, HDGS51654, HDGS51653, HDGS51652, HDGS51651, HDGS51650, HDGS51649, HDGS51648, HDGS51647, HDGS51646, HDGS51645, HDGM51179, HDGQ50425, HDGN50465, HDGN50849, HDGN50467, HDGN50470, HDGN50475, HDGR52324, HDGR52323, HDGR52322, HDGR52321, HDGN50468. f. Model Number M1145476; No GTIN Number; Serial Numbers: HDGR52896. g. Model Number M1235957; No GTIN Number; Serial Numbers: HDGT60222. h. Model Number M1149051; No GTIN Number; Serial Numbers: HDGR53018, HDGR53085, HDGR52927, HDGR53086, HDGR53021. i. Model Number M1146098; No GTIN Number; Serial Numbers: HDGS52233, HDGS52234.

Root Cause

Nonconforming Material/Component

Action Taken

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees beginning on 05/19/2023 by mail. The notice explained the problem, risk, and requeested the following: Actions to be taken by Customer/User: Inspect the device bedside panels; if the device has the correct latches, continue to use the device. If the device does not have the correct latches, quarantine the device. Inspect any field replaceable bedside panel stock and quarantine for the affected product. GE Healthcare will replace affected latches.