Irvine Scientific Oil for Embryo Culture (Light Mineral Oil) REF Catalog # 9305

Company
Fujifilm Irvine Scientific, Inc.
Recall Initiated
January 16, 2023
Posted
April 1, 2023
Recall Number
Z-1314-2023
Quantity
15,131 bottles
Firm Location
Santa Ana, CA

Reason for Recall

Due to the detection of oil toxicity in several lots of Oil for Embryo Culture products.

Distribution

U.S.: AZ, AR, CA, CT, DC, FL, HI, IA, ID, IL, IN, KS, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WI and WV. O.U.S.: India, Israel, Japan, Malaysia, Mexico, South Korea, South Africa, Thailand, Vietnam, Brazil, Bosnia, Argentina, Netherlands, and United Kingdom

Lot / Code Info

UDI-Di Code: 08937270029305TH Catalog Number / Lot Numbers: 9305-100ml / 0000011351, 9305-500ml / 0000011367, 9305-100ml / 0000015999, 9305-500ml / 0000016001

Root Cause

Nonconforming Material/Component

Action Taken

On 01/16/2023, the firm emailed an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that, the firm has received complaints of reduced blastocyst development, including complete degradation in some cases, during culture of human embryos. Additionally, some clinics performed Human Sperm Survival Assay (HSSA), as part of incoming confirmatory Quality Control testing, and reported low sperm motility. Customers are instructed to: 1. Immediately stop using the product; 2. Quarantine any remaining product in inventory; 3. Return or destroy the product; 4. Complete and return the enclosed MEDICAL DEVICE RETURN RESPONSE FORM; and 5. Email the completed MEDICAL DEVICE RETURN RESPONSE FORM to fisira@fujifilm.com by January 23, 2023. For questions, contact: Customer Service Manager, Medical Business: - Toll free in the US: 1.800.577.6097 - Phone: 1.949.261.7800 x457 - Email: Melissa Gomez, fisitmrequest@fujifilm.com