RecallDepth

Optima Coil System Model Number OPTI0208CSS10, Model Number OPTI0308CSS10. Used in endovascular procedures.

Company
Balt Usa, Llc
Recall Initiated
September 7, 2022
Posted
October 20, 2022
Recall Number
Z-0103-2023
Quantity
41 impacted devices
Firm Location
Irvine, CA

Reason for Recall

Product pouch label does not match up with carton label

Distribution

Worldwide distribution - U.S. Nationwide distribution in the states of FL, GA, NC, and PA. The countries of Dominican Republic, India and Japan.

Lot / Code Info

Model Number: OPTI0208CSS10 UDI-DI Code: 00818053025815 Lot Number: F220601068 Model Number: OPTI0308CSS10 UDI-DI Code: 00818053025891 Lot Number: F220601089

Root Cause

Error in labeling

Action Taken

On 09/07/2022, the firm sent an "URGENT MEDICAL DEVICE RECALL" Letter via priority mail to customer informing them that Balt USA was informed that one device had information on the pouch label that differed from the information on the carton label. This hazard presents a high risk to patient safety with the possibility that the user could inadvertently select the wrong size coil for a procedure. Distributor/Subsidiaries are instructed to: -Inform customers about this notice. -Return to Balt USA products from Lot #F2206001068 and Lot #F220601089 Direct Customers are instructed to: -Inform within their hospital Safety Office, Pharmacists, Head of Neuroradiology and staff of Recall. -Return to Balt USA products from Lot #F2206001068 and Lot #F220601089 For questions or assistance - contact Balt USA Quality Department at email - QA@baltgroup.com telephone number - 949-788-1443 or Fax 949-788-1444

More recalls by Balt Usa, Llc

View all recalls by this company →