Medisafe Distal Duck Kit and Duck Bag, Cleaner, ultrasonic, medical instrument - Product Usage: intended to keep instruments moist during transportation between point of use and cleaning as an initial step of reprocessing.

Company
Steris Corporation
Recall Initiated
November 19, 2020
Terminated
October 25, 2023
Recall Number
Z-0830-2021
Quantity
23,445 units, Domestic 5,727 units, Foreign 17,718 units
Firm Location
Mentor, OH

Reason for Recall

STERIS has identified that certain lots of diluted 4-Zyme may contain bacteria, specifically Pseudomonas fluorescens. The presence of this bacteria can cause the color of the detergent to darken over time. There is an improbable risk to users of the product from exposure to this bacteria, and no risk to patients.

Distribution

Worldwide distribution - US Nationwide distribution including in the states of CA, DE, CO, VA, FL, WA, TX, OH, PA, MA, IA, CT and the countries of United Kingdom, Saudi Arabia, Portugal, Denmark, Ireland, United Arab Emirates, South Korea, Australia, Greece, Ashmore & Cartier Islands, Belgium, Canada, China, Germany, Japan, Norway, Poland, Qatar, Romania, Sweden, Thailand.

Lot / Code Info

Model# M20350 Lot# 1811273,1811277,1906484,1908593,1909635,1910651, 1911658,2001730,2004770  Model# M20358 Lot# 1903408,1906508,1907556,1909644,2001718,2004772  Model# M20359 Lot# 1811278,1906499,1910649,1911680,1911691,1912708, 2001719,2004768  Model# M20400 Lot# 1901340,1901349,1902374,1904448,1904464,1905505, 1906529,1907571,1908595,1909624,1910653,1912700, 2001717,2004766,2004774,2006797

Root Cause

Device Design

Action Taken

On 11/23/2020, STERIS issued Urgent Device Correction notices to customers via courier service. Instructing customers to inspect and destroy any remaining affected products in their inventory. STERIS will provide replacement product at no cost to Customers who have product remaining in stock.