TactiCath Quartz Contact Force Ablation Catheter, Model Numbers: PN-004 065 (GTIN 07640157990033) and PN-004 075 (GTIN 07640157990040) - Product Usage: is a multi-electrode irrigated catheter with a deflectable tip designed to allow electrophysiological mapping of the heart and, when connected to an RF generator, to deliver radiofrequency current to the catheter tip electrode for the purpose of ablation.

Company
Abbott
Recall Initiated
May 22, 2020
Terminated
February 11, 2021
Recall Number
Z-2541-2020
Quantity
84104 devices
Firm Location
Plymouth, MN

Reason for Recall

High rate of reported complaints for the loss of contact force information while using the TactiCath Quartz Contact Force Ablation Catheter

Distribution

Worldwide distribution.

Lot / Code Info

All batch numbers with an expiration date of April 25, 2020

Root Cause

Process control

Action Taken

The firm began notifying their consignees by letter on 05/22/2020. The notification reported the high rate of reported complaints for loss of contact force information while using the product, the risk to patient, and recommendations to avoid contact force failures.