Elekta Unity, Image-Guided Radiation Therapy System

Company
Elekta, Inc.
Recall Initiated
December 18, 2019
Terminated
June 15, 2021
Recall Number
Z-2473-2020
Quantity
21 devices
Firm Location
Atlanta, GA

Reason for Recall

There have been reports of excessive heating in the MRL Gradient Connection Module.

Distribution

Distributed in US - PA, NY, and IA. Distributed ROW - Australia, Canada, China, Switzerland, Denmark, Hong Kong, Italy, Netherlands, Sweden, Thailand, Turkey, UK. There was no government/military distribution.

Lot / Code Info

Elekta Unity Serial numbers 600014, 600016, 600017, 600018, 600019, 600020, 600021, 600022, 600023, 600024, 600025, 600026, 600027, 600028, 600029, 600030, 600031, 600032, 600033, 600034, 600035

Root Cause

Process control

Action Taken

Initially, the firm issued an email on 12/18/2019 to their field engineers which said they had attached a "Knowledge Article" which provided instructions on how to inspect the Philips Marlin 1.5T system used in the Elekta Unity MR Linac. On 5/11/2020, the firm issued an Urgent Important Field Safety Notification dated May, 2020, via email to consignees which explained the problem, clinical impact, and said that initial corrective actions have been performed on all impacted sites to ensure the gradient coil connection assembly is correct. It goes onto say that extensive testing of the incorrect connection material has been completed to ensure that continued use of the Elekta Unity system is acceptable, however, it remained that further actions can be taken to reduce the risk.