Professional Urine Drug Control, Item code 88010 - Product Usage: The Alere Professional Cup Urine Drug Controls are compatible with all quantitative and qualitative drug detection procedures which are sufficiently sensitive to detect the control constituents.
- Company
- Cardinal Health Inc.
- Recall Initiated
- April 20, 2020
- Terminated
- July 23, 2020
- Recall Number
- Z-2050-2020
- Firm Location
- Dublin, OH
Reason for Recall
Three products requiring refrigerated storage conditions were incorrectly stored outside of approved temperature ranges. Using these temperature abused products may result in misdiagnosis of the patient.
Distribution
US Nationwide distribution NC, KS, and AL.
Lot / Code Info
Lot KN00146
Root Cause
Storage
Action Taken
On April 20, 2020, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were informed that the product had been incorrectly stored and that there was a potential for misdiagnosis. Action Required: 1. CHECK all storage and usage locations to confirm whether you have any units of the affected product codes and lot numbers containing the labeling outlined in the letter. 2. SEGREGATE and QUARANTINE all on-hand affected product. 3. PLEASE RETURN the letter's acknowledgment form via facsimile (847-689-9101 or 614-652-9648) or email (GMB-FieldCorrectiveAction@cardinalhealth.com) and indicate the product code, lot and quantity of product youve quarantined or discarded. Please respond regardless of whether or not you have affected product. 4. NOTIFY any customers to whom you may have distributed, or forwarded product affected by this recall. Your notification to your customers may be enhanced by including a copy of this recall notification letter. 5. CONTACT the appropriate Customer Service group to arrange for return and credit/replacement of any affected product Monday Friday between 8:00am - 5:00pm EST: " Hospital800-964-5227 " Federal Government800-444-1166 " Distributor800-635-6021 " All other Customers888-444-5440