Oxymetry Cable, Rxonly. REF/UDI: HEMOXSC100/(01)00690103196986 - Product Usage: The HemoSphere oximetry cable is a multi-use device that connects with a compatible Edwards monitor on one end and any approved Edwards oximetry catheter on the other end to continuously measure venous oxygen saturation.

This recall has been terminated (originally issued May 8, 2020).

Company
Edwards Lifesciences, Llc
Recall Initiated
October 7, 2019
Posted
May 8, 2020
Terminated
April 27, 2021
Recall Number
Z-1928-2020
Quantity
35
Firm Location
Irvine, CA
Official Source
View on FDA website ↗

Reason for Recall

Intermittent communication due to inconsistent crimping of connectors onto cables.

Distribution

Worldwide distribution - US Nationwide distribution including in the states of AL, CA, FL, IL, IN and MO, and the countries of Germany, Netherlands, United Kingdom.

Lot / Code Info

The cables with the following serial numbers (35) are recalled: 13710131, 13710130, 13710129, 13710115, 13710113, 13710025, 13710027, 13710028, 13710029, 13710030, 13710031, 13710041, 13710043, 13710045, 13710050, 13710051, 13710056, 13710057, 13710058, 13710061, 13710062, 13710063, 13710064, 13710066, 13710069, 13710070, 13710084, 13710086, 13710089, 13710092, 13710093, 13710094, 13710098, 13710099, 13710100.

Root Cause

Process control

Action Taken

On xx/xx/2019, and URGENT MEDICAL DEVICE RECALL-ACTION REQUIRED, letter was sent to customers by FedEx. Customers are instructed to acknowledge this notification , contact the firm at the contact information in the communication to obtain a Returned Goods Authorization (RGA) number and to return all product to recalling firm.