RX Daytona, Model Nos. RX4040 and RX4041

This recall has been terminated (originally issued June 24, 2019).

Company
Randox Laboratories, Limited Ardmore; 55 The Diamond Road Crumlin United Kingdom
Recall Initiated
June 24, 2019
Terminated
May 21, 2020
Recall Number
Z-2083-2019
Quantity
3
Official Source
View on FDA website ↗

Reason for Recall

There was a report of a malfunction with the sensor which resulted in an increase of temperature leading to melting of internal wires and scorch damage to the internal structure of the instrument.

Distribution

Domestic distribution to CA CT MA MD MO NC NJ NY OH PA UT WV and PR. Worldwide foreign distribution.

Lot / Code Info

All serial numbers

Root Cause

Under Investigation by firm

Action Taken

On June 24, 2019, the firm notified its customers of the recall through an Urgent Field Safety Notice. Customers were asked to take the following actions: If there is any sign of damage to the RCU lid or lid sensor " Switch of main analyser main power supply as soon as possible " Contact technical support immediately. " Discuss the contents of this notice with your Medical Director. " Display this notice in a prominent location near the instrument. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days.