Talar Dome Total Ankle Prosthesis , Sloped, Size 0, Right, Product No. LJV180T
- Company
- Integra Lifesciences Corp.
- Recall Initiated
- March 8, 2019
- Terminated
- May 15, 2020
- Recall Number
- Z-1168-2019
- Quantity
- 2029
- Firm Location
- Plainsboro, NJ
Reason for Recall
The Salto Talaris and Integra XT Revision implant labelling was found to list an incorrect metallic coating material symbol on the metallic implant labels (box and inner pouch) and IFU
Distribution
Distributors in AL, CA, CT, FL, GA, HI, IA, ID, KS, KY, MD, MI, MO, MT, NC, NE, NJ, NM, NY, OH, OK,PA, TX, VA, WA, and WI.
Lot / Code Info
182853 182854 183344
Root Cause
Other
Action Taken
On March 8, 2019, the firm notified affected customers and distributors of the product issue via Urgent Medical Device Correction letter. The letter advised customers that the incorrect metallic coating material symbol had been printed on the implants labeling. Customers were asked to do the following: Please examine your inventory to determine if you have any Salto Talaris Total Ankle Prosthesis implants and/or Integra XT Revision Total Ankle Replacement System implants. Customers: 1. If you do have any impacted lots of Salto Talaris Total Ankle Prosthesis implants and/or Integra XT Total If you have any impacted lots of Salto Talaris Total Ankle Prosthesis implants and/or Integra XT Total Revision Ankle Replacement System implants, please be aware the correct material symbol is CP Ti. The IFU containing the correct material symbol can be accessed at the link below. Product removal is not required. Distributors: I1. f you have any impacted lots of Salto Talaris Total Ankle Prosthesis implants and/or Integra XT Total Revision Ankle Replacement System implants, please be aware the correct material symbol is CP Ti. The IFU containing the correct material symbol can be accessed at the link below. Product removal is not required. "2. We also kindly ask that you provide this letter to impacted customers who may have affected products. IFU Link: Integra XT: https://www.integralife.com/file/general/1551465297.pdf Salto Talaris: https://www.integralife.com/file/general/1546543287.pdf Customers and distributors were asked to complete the attached Acknowledgement Form and return the completed form by email or fax indicating receipt and review of this notification. Keep a copy of the form for your records. Should you have any questions regarding these instructions, please contact Customer Service Monday to Friday 8:00 AM 8:00 PM EST USA: 1-800-654-2873: email: custsvcnj@integralife.com