JS Series SterilContainer S2 System
- Company
- Aesculap Implant Systems Llc
- Recall Initiated
- September 8, 2017
- Posted
- February 15, 2018
- Terminated
- April 4, 2019
- Recall Number
- Z-0624-2018
- Quantity
- 292 units
- Firm Location
- Center Valley, PA
Reason for Recall
The JS Series SterilContainer S2 System was released to the market place prior to an authorized FDA 510K clearance. No other Aesculap SterilContainer System products are impacted by this recall.
Distribution
Nationally
Lot / Code Info
JS440 S2 1/1CONT.BOTTON PERFOR.H88MM H202 JS441 S2 1/1 CONT.BOTTOMPERFOR.H1118MM H202 JS442 S2 1/1-CONT.BOTTOM PERFOR.H133MM H202 JS444 S2 1/1 CONT.BOTTOM PEFOR.H185MM H202 JS489 S2 1/1-CONTAINER LID H202 JS740 S2 3/4 CONT.BOTTOM PERFOR.H88MMH202 JS741 S2 3/4 CONT.BOTTOM PERFOR.H118MM H202 JS742 S2 3/4 CONT.BOTTOM PERFOR.H133MM H202 JS789 S2 3/4 CONTAINER LID H202 JS340 S2 1/2 CONT.BOTTOM PERFOR.H88MM H202 JS341 S2 1/2 CONT.BOTTOM PERFOR.H118MM H202 JS342 S2 1/2 CONT.BOTTOM PERFOR.H133MM H202 JS389 S2 1/2 CONTAINER LID H202
Root Cause
No Marketing Application
Action Taken
On August 24, 2017 a Medical Device Recall(Reference 2916714-08/24/2017-010R) was issued requesting that all customers take the following steps: 1.Review the Urgent Medical Device Recall Notification Letter in its entirety and ensure that all users in your organization and other concerned persons are informed about this recall and immediately discontinue use. 2. Determine your current inventory of affected containers within your facility. Immediately quarantine the affected products. A sales representative will come to your facility and remove your affected product and immediately return it to Aesculap Inc. They will assist you with ordering an appropriate replacement container. Do not destroy any affected products. 3. Please contact Customer Service and they will provide you a Returned Good number and a return label. Customer Service: Phone: 1-800-282-9000 aesculap_emailorders.bbmus_service@bbraun.com