Brilliance iCT Intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes.

This recall has been terminated (originally issued November 10, 2017).

Company
Philips Medical Systems (Cleveland) Inc
Recall Initiated
November 10, 2017
Terminated
November 5, 2020
Recall Number
Z-0388-2018
Quantity
13
Firm Location
Cleveland, OH
Official Source
View on FDA website ↗

Reason for Recall

Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the fasteners may become loose, which may have a potential for harm.

Distribution

Worldwide Distribution -- US, Argentina, China, Republic of Korea, Russian Federation, Slovakia, Turkey, and United Kingdom. 4 in US, (1 military), 14 foreign accounts

Lot / Code Info

Code 728306, SN# 100776, UDI: (01)00884838059474(21)100776; 100778, UDI: (01)00884838059474(21)100778; 100779, UDI: (01)00884838059474(21)100779; 100780, UDI: (01)00884838059474(21)100780; 100781, UDI: (01)00884838059474(21)100781; 100782, UDI: (01)00884838059474(21)100782; 100783, UDI: (01)00884838059474(21)100783; 100784, UDI: (01)00884838059474(21)100784; 100785, UDI: (01)00884838059474(21)100785; 100787, UDI: (01)00884838059474(21)100787; 100790, UDI: (01)00884838059474(21)100790; 100791, UDI: (01)00884838059474(21)100791; and  100794, UDI: (01)00884838059474(21)1007.

Root Cause

Under Investigation by firm

Action Taken

An Urgent - Field Safety Notice Medical Device Correction letter, dated November 1, 2017, was sent to customers. The letter identified the affected device, described the issue, and discussed the hazard involved. Users are to cease use immediately and contact their Philips Field Service Engineer, if any unusual noises are heard within the gantry, or if the patient support tabletop is in free float. Philips Field Service Engineers will correct the system by replacing the affected fasteners. If further info or support concerning this issue is needed, customers can contact their local Philips rep or the Customer Care Solutions Center at 1-800-722-9377.