RecallDepth

Bigger-Better-Bladder With 3/8" ID tubing, ITEM BBB38

Company
Circulatory Technology Inc
Recall Initiated
March 3, 2016
Terminated
April 20, 2017
Recall Number
Z-2050-2016
Quantity
72 units
Firm Location
Oyster Bay, NY

Reason for Recall

The seal between the balloon and the housing may leak.

Distribution

US Distribution to the states of: CA, IN, MI, NY, PA, TX & VA

Lot / Code Info

Lot # 014603 USE BY: 9/2018

Root Cause

Device Design

Action Taken

Circulatory Technology Inc. sent an Urgent Product Recall notices dated March 3, 2016, to their customers via email. Circulatory Technology Inc. has identified this action as a voluntary recall, however they did not withdraw the product from the market. Instead, Circulatory Technology Inc. has advised customers to check the integrity of the seal between both the inlet and outlet tube and the housing as per the Instructions For Use. In addition, for this lot, Circulatory Technology Inc. recommends that you bend the inlet as well as the outlet tube away from the housing before conducting the leak test. There should be no visible separation between the tubing and the housing. Customers with questions can call 516-624-2424.

More recalls by Circulatory Technology Inc

View all recalls by this company →