OMNIlife science, Inc. ApeX acetabular liner hip prosthesis component Serf Dual Mobility Cup Insert Liner 51-28; Product code: H6-25128

This recall has been terminated (originally issued July 23, 2019).

Company
Omnilife Science Inc.
Recall Initiated
July 23, 2019
Terminated
April 24, 2020
Recall Number
Z-2480-2019
Quantity
22 units
Firm Location
Raynham, MA
Official Source
View on FDA website ↗

Reason for Recall

Acetabular liners failed the oxidation testing performed prior to release. They may experience a higher wear rate, leading to revision surgery.

Distribution

US nationwide distribution to states of: IL, NY, OK, MD, AL, PA, CO, FL, and UT.

Lot / Code Info

LOT # 31740

Root Cause

Nonconforming Material/Component

Action Taken

OMNIlife science, Inc. sent an Urgent Medical Device Recall letter, dated July 23rd, 2019. 1. Identify the affected product from inventory and hold (quarantine) the product. 2. Return product to Omni at 480 paramount drive, Raynham, MA 02767 3. Acknowledge this notice by retuning the acknowledgement and receipt form by fax to 508-819-3390 or email to Omni-cs@coringroup.com For further information contact Christina Rovaldi, Regulatory Affairs Manager at 774-226-1847 or 800-448-6664. Return the attached Acknowledgement form by FAX to 508-819-3390 or by email to Omnics@coringroup.com. Company contact hours are Monday through Friday, 8:00 AM to 4:30 PM, Eastern Time.