Carina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 60600009
This recall has been terminated (originally issued July 28, 2016).
- Company
- Handicare Ab Maskinvagen
- Recall Initiated
- July 28, 2016
- Terminated
- December 17, 2018
- Recall Number
- Z-0014-2018
- Quantity
- 9 units
- Official Source
- View on FDA website ↗
Reason for Recall
A limited amount of Carina mobile lifts have been delivered with a too small locking pin. If a locking pin smaller than the correct size is mounted on the Carina mobile lift, the locking pin is likely to cause increased wear on the plastic bushings. In case of continued use, these components may cause wear on the metal of the piston rod eye.
Distribution
OH, PA, MD
Lot / Code Info
a) Carina350EM, REF 6060011, Serial numbers: 661102000 - 661102160 b) Carina350EML, REF 60600012, Serial numbers: 661200500 - 661200584 c) Carina350EE, REF 60600009, Serial numbers: 660900500 -660900571
Root Cause
Process control
Action Taken
The firm initiated their recall by email on 07/28/2017. The firm directed the consignee to repair the units on site and destroy the affected parts.