Medtronic Curved Suction 70 9733450 EM ENT, reusable, Rx. The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical tool has a set of receiving coils to sense the magnetic field strength. Using a mathematical algorithm, the position and orientation of the instrument can be computed and overlaid/registered to imagery on a computer monitor, resulting in the ability to track the instrument tip in relation to the patients anatomy. AxiEM ENT Suction devices are used as suction and pointing devices. The suction devices are intended to have an ENT Instrument Tracker mounted on it

This recall has been terminated (originally issued June 29, 2017).

Company
Medtronic Navigation, Inc.
Recall Initiated
June 29, 2017
Terminated
August 3, 2018
Recall Number
Z-3052-2017
Quantity
149 instruments
Firm Location
Louisville, CO
Official Source
View on FDA website ↗

Reason for Recall

The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.

Distribution

Worldwide Distribution - US (nationwide) and Internationally to Canada.

Lot / Code Info

Lot numbers 17030722, 17022421,17022020, 17020619, 17011918, 17012417, I 17010602, 16122724,16122239, 16122038, 16120563, and 16112362

Root Cause

Device Design

Action Taken

Medtronic sent an Urgent Medical Device Recall letter dated June 25, 2017, via overnight mail on June 29, 2017, to all affected customers. Letters to foreign customers are being handled by geography-specific Regulatory Specialists. Customers were instructed examine their inventory for the affected product and immediately quarantine them for return for Medtronic. Customers with affected product were instructed to contact Medtronic Technical Services at 1-800-595-9709 to receive a return material authorization (RMA) and arrange for their return and no charge replacements. Customers were asked to complete and follow instructions on the attached Consignee Response Form. Sign and date the bottom of the form and then return the form to Medtronic at RS NavFCA@medtronic.com or fax it to Medtronic Technical Services at 651-367-7075. Customers with questions were instructed to call 1-800-595-9709.