RecallDepth

Automated peritoneal dialysis (APD) cycler

This recall has been terminated (originally issued February 16, 2016).

Company
Baxter Healthcare Corp.
Recall Initiated
February 16, 2016
Terminated
December 23, 2016
Recall Number
Z-1565-2016
Quantity
10 units
Firm Location
Deerfield, IL
Official Source
View on FDA website ↗

Reason for Recall

Baxter Healthcare Corporation is sending this communication to inform you of incomplete instructions in the AMIA Automated Peritoneal Dialysis (PO) System Clinician Guide for calculating the total recommended solution therapy volume. Specifically, the instructions do not specify the need for an extra 200 mL of PD solution in order to prime the patient line and for air purge operations. The tot

Distribution

US Consignee: NY ***Foreign Consignee: Ecuador

Root Cause

Under Investigation by firm

Action Taken

An Urgent Device Correction communication was sent to the one affected customer via U.S.P.S., first class mail on 02/11/2016. A Clinician Guide Addendum was created to instruct clinicians of the extra 200 mL of PD solution required to prime the patient line and for air purge operations. This Addendum was provided as an attachment to the Urgent Device Correction communication letter.

More recalls by Baxter Healthcare Corp.

View all recalls by this company →