RecallDepth

Capnostream20 (US) M W/PRINTER REFURBISHED, Part Number CS78659

This recall has been terminated (originally issued July 31, 2018).

Company
Oridion Medical 1987 Ltd. Har Hotzvim Industrial
Recall Initiated
July 31, 2018
Terminated
September 20, 2022
Recall Number
Z-0035-2019
Quantity
22
Official Source
View on FDA website ↗

Reason for Recall

The date/time, nurse call and alarm settings of the bedside patient monitors may reset to the factory default settings when the monitor is powered off.

Distribution

The products were distributed US nationwide and Puerto Rico. The products were distributed to the following foreign countries: Argentina, Australia, Austria, Bahrain, Belgium, Canada, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Kazakhstan, Kuwait, Mexico, Netherlands, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, UAE, UK, and Vietnam.

Lot / Code Info

B200001278 B300001166 B300001220 B300001070 B300001092 B300001097 B300001108 B300001172 B300001181 B300001185 B300001205 B300001278 B300001115 B500001885 B500001831 B300001015 B300001118 B500002336 B300001012 B300001132 B300001139 B500001819

Root Cause

Nonconforming Material/Component

Action Taken

Urgent Medical Device Correction notification letters dated 7/31/18 were distributed to customers. The letters included the following instructions for customers: Share this notification with all care environments where the Capnostream" 20 and Capnostream" 20p bedside patient monitors are used, particularly in the areas where nurse call may be enabled (general care floors). If your facility has distributed these bedside patient monitors to other persons or facilities, please promptly forward a copy of this letter to those recipients. Medtronic is recommending continued use of Capnostream" devices. Users should confirm that the date/time is accurately displayed. If the date/time requires reset, all user defined settings, such as alarms and nurse call, will also require reset until the software is updated. Please follow the instruction in the Operators Manual to set user-defined default settings. The Operators Manual is available at http://www.medtronic.com/covidien/enus/support/product-manuals.html, refer to the Institutional Settings section. Return the completed Acknowledgement and Receipt Form by fax or email even if you have no inventory. In October 2018, download the software update from the Medtronic website via the following link which includes directions on how to download and install the software: http://www.medtronic.com/covidien/en-us/support/patient-monitoring-equipmentsoftware-upgrades.html.

More recalls by Oridion Medical 1987 Ltd. Har Hotzvim Industrial

View all recalls by this company →