MRSASelect II, IVD, REF 63758 (20 plates per package). The MRSASelect is indicated for use in conjunction with other laboratory tests and clinical data available to aid in the identification and diagnosis of MRSA from patients with skin and soft-tissue infections.

Company
Bio-Rad Laboratories, Inc
Recall Initiated
June 8, 2018
Terminated
October 30, 2020
Recall Number
Z-1743-2019
Quantity
3,632 units
Firm Location
Redmond, WA

Reason for Recall

Abnormal coloration of agar plates and the growth of non-characteristic MRSA colonies.

Distribution

US Nationwide Distribution to: AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, and Puerto Rico.

Lot / Code Info

Lot numbers 64177778, exp. 6/10/2018; 64179165, exp. 6/17/2018; 64182547, exp. 7/3/2018; 64184389, exp. 7/8/2018; 64187554, exp. 7/22/2018; 64189124, exp. 7/23/2018; and 64190805, exp. 7/31/2018.

Root Cause

Nonconforming Material/Component

Action Taken

The firm. Bio-Rad laboratories, Inc., issued an "URGENT PRODUCT CORRECTION" letter dated 6/7/2018 to all customers via FedEx on 6/7-8/2019. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: " As already recommended in the package insert, please ensure that exposure of agar plates to light is minimized, both before and during incubation, with the added time restriction of no more than 2 hours during storage, warming, incubation, and reading. " Do not use agar with a change in color from white to an orange-pinkish shade. " As recommended in the package insert, if in doubt, confirm the identification of colonies by coagulase test or PASTOREX STAPH-PLUS. Please feel free to contact Technical Support with any questions or concerns at 1-800-224-6723, option #2 and then #3.