Adapter for handpiece, guided, reusable, for guide sleeve 0 4.8 mm Article number: 3.04.090 - Product Usage: The adapter for handpiece, guided reusable, for guide sleeve 4.8 mm is used to transfer the implant from the packaging to the patient via a plug-in connection and screwed in manually or mechanically.

This recall has been terminated (originally issued April 3, 2019).

Company
Thommen Medical Ag Eschenweg 3 Bettlach Switzerland
Recall Initiated
April 3, 2019
Terminated
August 12, 2019
Recall Number
Z-1390-2019
Quantity
15 US
Official Source
View on FDA website ↗

Reason for Recall

Dental adaptor does not conform to specifications. The dental coupling of the adapter can become deformed when being inserted into the implant may make it difficult to remove the adapter from its counterpart (e.g. the handpiece).

Distribution

US Nationwide distribution in the state of OH.

Lot / Code Info

Lot number:17415 UDI: 07640182640705

Root Cause

Under Investigation by firm

Action Taken

Thommen Medical AG notified the account on 3/27/19 via email asked distributors to put any adapters in their stock in quarantine. Furthermore recall all adapters shipped to their customers. All recovered adapters to be sent back to Thommen Medical AG, who will supply them with new adapters which conform to the specifications.