RecallDepth

Heater Cooler Unit Usage: The Heater-Cooler Unit (HCU 30) supplies temperature-controlled water for cardioplegia heat exchangers and for blood heat exchangers in extra corporeal circulation and for blankets with which patients can be warmed or cooled. The HCU 30 is intended to circulate water through heat exchange circuits to warm or cool a patient during short duration cardiopulmonary bypass procedures lasting 6 hours or less.

This recall has been terminated (originally issued September 27, 2016).

Company
Maquet Cardiovascular Us Sales, Llc
Recall Initiated
September 27, 2016
Terminated
August 19, 2019
Recall Number
Z-0348-2017
Quantity
305 units
Firm Location
Wayne, NJ
Official Source
View on FDA website ↗

Reason for Recall

There is the potential for contaminated water to enter parts of the Heater-Cooler Unit circuits. There is a potential risk of bacterial contamination including Mycobacteria species.

Distribution

Worldwide Distribution - US Nationwide

Lot / Code Info

HCU 30 Base Unit 100-120 V, Model Number 704629 I 70103.4371 HCU 30 Base Unit Canadian/US, Model Number 705501 I 70103.4653 HCU 30 Base Unit 200-240 V, Model Number 939001 I 70102.8718 All serial numbers are affected.

Root Cause

Device Design

Action Taken

Maquet issued an Urgent Medical Device Recall letter/Return Response form dated September 27, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. Customer were instructed to complete the return response form via fax to 1-973-629-1518. For questions contact Maquet Customer Service at 1-888-627-8383 option 2 followed by option 2 Monday through Friday between 8 amd and 6 pm. EST.