RecallDepth

VITROS Immunodiagnostic Products Estradiol Reagent Pack, Unique Device Identifier No. 10758750005017, REF/Catalog Number/Product Code 855 2630; IVD -- Ortho-Clinical Diagnostics, Pencoed, Bridgend CF35 5PZ, UK --- The VITROS Estradiol test is performed using the VITROS Estradiol Reagent Pack and Calibrators on VITROS ECi/ECiQ Immunodiagnostic Systems, VITROS 3600 Immunodiagnostic Systems and VITROS 5600 Integrated Systems.

This recall has been terminated (originally issued July 14, 2016).

Company
Ortho-Clinical Diagnostics Felindre Meadows Bridgend United Kingdom
Recall Initiated
July 14, 2016
Terminated
June 8, 2018
Recall Number
Z-2352-2016
Quantity
Domestic: 1620; Foreign: 7597
Official Source
View on FDA website ↗

Reason for Recall

Ortho Clinical Diagnostics has issued a recall of their VITROS immunodiagnostics Estradiol Reagent assay. The firm was made aware of a potential cross-reactivity between the drug Fulvestrant and theVITROS Immunodiagnostic Estradiol Reagent assay. The firm conducted an internal investigation and was able to confirm biased positive (ranging from 1303-4046%) on Estradiol results, obtained from postmenopausal females taking Fulvestrant.

Distribution

Nationwide, Puerto Rico, and foreign distribution to the following countries: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Poland, Singapore, Spain, Venezuela.

Lot / Code Info

Lot numbers: 1470 (expiry date 02-Aug-16); 1480 (expiry date 05-Aug-16), 1490 (expiry date 10-Oct-16), 1500 (expiry date 10-Oct-16), 1510 (expiry date 01-Nov-16), 1528 (expiry date 01-Nov-16) 1538 (expiry date 14-Mar-17), 1548 (expiry date 09-Mar-17), 1558 (expiry date 09-Mar-17) and 1568 (expiry date 29-Mar-17)

Root Cause

Device Design

Action Taken

On 7/14/2016, URGENT PRODUCT CORRECTION NOTIFICATION Customer Letter (Ref. CL2016-141, dated 7/14/2016) was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all customers who received product since 6/1/2015 to inform them that Fulvestrant is an interferent of the Estradiol Reagent Pack and causes positively biased results. Discuss any concerns you may have regarding previously reported estradiol results with your Laboratory Medical Director to determine the appropriate course of action. The results from this or any other diagnostic test should be used and interpreted only within the context of the overall clinical picture. For questions, contact Customer Technical Services at 1-800-421-3311. Foreign affiliates were informed of the issue via e-mail on 7/14/2016 and instructed to notify their consignees of the issue.

More recalls by Ortho-Clinical Diagnostics Felindre Meadows Bridgend United Kingdom

View all recalls by this company →