Miltex Premium Grade Self-Lock Chisel Blade Handle, Catalogue No. 4-405, manual surgical instrument.

This recall has been terminated (originally issued May 31, 2016).

Company
Integra York Pa, Inc
Recall Initiated
March 11, 2016
Posted
May 31, 2016
Terminated
April 17, 2017
Recall Number
Z-1841-2016
Quantity
678 Chisel Blade Holders
Firm Location
York, PA
Official Source
View on FDA website ↗

Reason for Recall

One lot of self-lock chisel blade handles have a wider slot that holds the chisel blade that may result in the inability to firmly secure the blade in the handle.

Distribution

Worldwide distribution. US nationwide, Mexico, and Korea.

Lot / Code Info

Catalogue No. 4-405, Lot # 97327

Root Cause

Nonconforming Material/Component

Action Taken

Integra mailed out an Urgent: Voluntary Medical Device Recall letter to customers informing them to evaluate their inventory for the affected product. If product was shipped to customers, they were asked to identify their customers and to stop shipping immediately. Integra requested that customers complete the attached form that was sent with the notification indicating whether or not they had the affected product on hand and customer service will contact them to provide a Return Authorization number for the product's return. Customers could reach Customer Service at 866-854-8300.