Firebird NXG Spinal Fixation System Set Screw Driver used to insert a set screw into a spinal fixation construct. Rx Only Provided non Sterile.

This recall has been terminated (originally issued June 1, 2016).

Company
Orthofix, Inc
Recall Initiated
April 27, 2016
Posted
June 1, 2016
Terminated
November 23, 2016
Recall Number
Z-1881-2016
Quantity
36 units
Firm Location
Lewisville, TX
Official Source
View on FDA website ↗

Reason for Recall

There is a possibility that the subject set screw driver could disengage from the torque limiting handle when the counter torque wrench is allowed to contact the counter torque handle.

Distribution

KS, NV, TX, CA, ID, FL, MA, MI, NJ, AZ, NH, SC, UT, WA

Lot / Code Info

All lots

Root Cause

Device Design

Action Taken

Affected distributors will be provided a written notification of field correction via USPS Certified Mail or FedEx on 4/27/16. Distributors will be additionally instructed to provide field correction notifcations to hospitals, surgeons, or other distributors that they may have further ditributed the product to.