CORDIS AVANTI + CATHETER SHEATH INTRODUCER SYSTEM The product includes a Catheter Sheath Introducer (CSI) and a vessel dilator. A Catheter Sheath Introducer facilitates percutaneous entry of an intravascular device. A vessel dilator facilitates the percutaneous entry of the CSI by forming an atraumatic transition from the skin through the subcutaneous tissue to the vessel.

This recall has been terminated (originally issued March 15, 2016).

Company
Cordis Corporation
Recall Initiated
December 31, 2015
Posted
March 15, 2016
Terminated
August 25, 2016
Recall Number
Z-1174-2016
Quantity
1,985 units
Firm Location
Miami Lakes, FL
Official Source
View on FDA website ↗

Reason for Recall

Incorrect cannula of the sheath introducer (smaller than intended).

Distribution

US Nationwide Distribution including states of:AL, AZ, CA, CO, CT, FL, GA, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, NH, NJ,NY, OH, OR, PA, RI, TN, TX, VA, WA, WV, and Hawaii.

Lot / Code Info

Catalog No. 504656X (GTIN: 20705032010389) Lot No. 17305580, Expiration date: 2018-07-31.

Root Cause

Mixed-up of materials/components

Action Taken

The firm, Cordis, sent an 'Urgent MEDICAL DEVICE CORRECTION (Field Safety Notice) dated 12/31/2015 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to do the following: 1) Read the Description of the problem section carefully to fully understand the issue involved. 2) Check your inventory to determine if you have any remaining affected product in your possession. Check all storage and usage locations. The purpose is to identify the product, not to remove the product. 3) Keep a copy of this notice with any affected product. 4) Review, complete, sign and return the enclosed Acknowledgement Form directly to Cordis by fax to (908) 429-8287 Alternate fax (904) 928-5077. number or scan and email to MiamiFRA@its.jnj.com. 5) Share this letter with anyone who needs to be informed in your facility and in any facility where potentially affected devices may have been transferred. 6) Maintain awareness of this notice until all affected product has been consumed. If vessel dilator to sheath incompatibility is experienced, you can report the incident through the standard complaint process (1-800-327-7714 Option 3). For any health care professional with medical concerns, please contact Cordis to speak to a clinician at 1-800-327-7714 Option 3, Monday through Friday from 9:00 AM to 5:00 PM EST. For questions related to the Correction that are not adequately addressed in this letter, please contact the Cordis Field Action Coordinator at (786) 313-2365. For questions related to the Acknowledgement Form and its return, please contact Cordis QA at (786) 313-8730."