Philips Healthcare TraumaDiagnost; Codes: 72001, 720011, 720013, 720020; A radiological examination unit for emergency work and routine skeletal diagnosis.

This recall has been terminated (originally issued July 22, 2015).

Company
Philips Electronics North America Corporation
Recall Initiated
July 22, 2015
Terminated
October 14, 2020
Recall Number
Z-0766-2016
Quantity
399 distributed, 262 are subject to this event
Firm Location
Andover, MA
Official Source
View on FDA website ↗

Reason for Recall

If arm is not fastened securely it could fall down.

Distribution

Distributed in the states of NY, WI, MA, AL, FL, GA, RI, CA, KS, and TX, and the countries of Canada, Australia, Austria, Belgium, France, Germany, Ireland, Italy, Netherlands, New Zealand, Portugal, South Africa, Spain, and Switzerland.

Lot / Code Info

All TraumaDiagnost products that were delivered before the week of July 2, 2007.

Root Cause

Nonconforming Material/Component

Action Taken

Consignees were notified of the recall via letter sent certified return receipt on 7/22/2015. The letter informs customers that the Instruction for Use (IFU) requires the operator not to use the system and call for service, if a mechanical defect or malfunction is suspected. In this case, it is recommended not to position the Ceiling suspension with the Trauma arm directly above a patient as long as this Field Change Order is not yet installed. Philips plans to repair the affected systems. A Philips Service Engineer will contact customers as soon as the Field Action Kit is ready to be implemented. For questions, contact your local representative or use the Technical Support Line at 1-800-722-9377. When communicating with Philips regarding this issue, please reference Field Change Order 72000014.