Airways Development LLC WaterPAP Positive Airway Pressure Device Reference #WP-7700.
This recall has been terminated (originally issued December 24, 2015).
- Company
- Airways Development Llc
- Recall Initiated
- September 24, 2015
- Posted
- December 24, 2015
- Terminated
- May 9, 2016
- Recall Number
- Z-0552-2016
- Quantity
- 1,464 units
- Firm Location
- Kenilworth, NJ
- Official Source
- View on FDA website ↗
Reason for Recall
Airways Development LLC has received a complaint about a canister leaking from WaterPAP Lot 8605A. During a review with their supplier it could not be determined if this was an isolated incident or if the whole lot is impacted.
Distribution
Nationwide
Lot / Code Info
Reference #WP-7700; Lot 8605A
Root Cause
Nonconforming Material/Component
Action Taken
Airways Development Inc. sent Product Alert notices dated 9/24/2015 to their customers via email. The alert informed customers about the related product issue and how to identify affected product. Airways Development is replacing any affected product left in their customers stock. Customers were also instructed to notify their customers if the product was further distributed. A point of contact was provided if the customers had any related questions.