Pinnacle3 Software Version 9.0, 9.2, 9.4 and 9.6, Model Numbers 453560446041, 459800091001, 459800220161, 459800232931, 459800235871, 459800338451. UPDATED: Pinnacle3 Software Version 8.0h, 8.0k, 8.0m, 8.0n.

This recall has been terminated (originally issued July 22, 2015).

Company
Philips Electronics North America Corporation
Recall Initiated
July 31, 2014
Posted
July 22, 2015
Terminated
May 8, 2019
Recall Number
Z-2200-2015
Quantity
1642
Firm Location
Andover, MA
Official Source
View on FDA website ↗

Reason for Recall

Philips, Pinnacle Radiation Treatment Planning System version 8.0h, 8.0k, 8.0m, 8.0n, 9 0, 9 2 9 4, 9 6 is being recalled because the dose may be inconsistent with the density of a density-overridden ROl.

Distribution

Worldwide Distribution-US: All states in the US including DC and PR. OUS: Australia, Austria, Bangladesh, Belgium, Brazil, China, Denmark, France, Germany, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Malaysia, Mexico, Netherlands,New Zealand, Oman, Philippines, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, United Kingdom, Viet Nam. UPDATED: Canada, Bulgaria, Egypt, Hungary, Jordan, Luxembourg, Poland, Reunion, Romania, Russia, Sauth Arabia, Turkey.

Lot / Code Info

453560446041, 459800091001, 459800220161, 459800232931, 459800235871, 459800338451  UPDATED:  453560446051 [A]  459800096361 [Al 453560446061 IA]  459800096541 [A] 453560446091 [A]  459800096741 [A] 453560446101 [A]  459800096921 [A] 453560446111 [A]  459800097111 [A] 453560446131 [A]  459800097291 [A] 453560446141 [A]  459800097471 [A] 453560446161 [A]  459800097651 [B] 453560446181 [A]  459800097831 [B] 453560446201 [A]  459800098171 [B] 459800089221 [A]  459800098351 [A] 459800091451 [A]  459800109261 [A] 459800138101 [BJ 459800189431 [A) 459800220131 [B] 459800239171 [A] 459800233271 [8] 459800233281 [A] 459800233291 [B] 459800233301 [A] 459800233311 [8] 459800233321 [B] 459800233331 [B] 459800233341 [A] 459800233351 [A] 459800233361 [A] 459800234101 [A] 459800234111 [A] 459800234121 [A] 459800318391 [Al 459800405751 [A] 459800417961 [A] 459800249651 [8] 459800249721 [A] 459800249851 [A] 459800249931 [A] 459800250001 [Al 459800250081 [A] 459800250151 [A] 459800250221 [A] 459800250291 [A] 459800250361 [A] 459800250431 [A] 459800250501 [A] 459800250571 [A] 459800338961 [B] 459800340181 [A] 459800340201 [A] 459800341171 [A] 459800341191 [A] 459800341481 [A] 459800341501 [A] 459800345001 [A] 459800345021 [A] 459800345041 [A] 459800345061 [A] 459800345081 [A] 459800345101 [A] 459800454251 [A] 459800455341 [A] 459800493591 [A] 459800494011 [A] 459800554831 [A] 459800554881 [A] 453560435761 453560434881 453560454791 453560461331 453560461341

Root Cause

Software design

Action Taken

Consignees were sent on 7/31/2014 a Philips "Urgent Field Safety Notice" dated 2014-July 20. The letter described the problem, product involved in the recall and the action that should be taken by the customers. For questions contact Customer Care Solutions Center at 1-800-722-9377. UPDATE: A second Philips letter dated 2 July 2015 was sent in August 6, 2015. This letter updated the recall to incorporate the additional products being recalled.