Winco Care Cliner X-Large - with Nylon Casters Oncology, Dialysis, Hospital Recovery, Home Dialysis Use, Physician Facilities.
This recall has been terminated (originally issued August 14, 2015).
- Company
- Winco Mfg., Llc
- Recall Initiated
- April 2, 2015
- Posted
- August 14, 2015
- Terminated
- December 14, 2016
- Recall Number
- Z-2393-2015
- Quantity
- 1 chair
- Firm Location
- Ocala, FL
- Official Source
- View on FDA website ↗
Reason for Recall
Shipment of foam for chairs was received classified as Fire Retardant but was in fact standard foam and would not meet Fire Retardant standards.
Distribution
Nationwide Distribution including CA, FL, KS, KY, LA, MA, MD, MN, MS, NJ, NY, OK, TN, TX, and WA.
Lot / Code Info
Model Number: 6540
Root Cause
Nonconforming Material/Component
Action Taken
Winco sent a field correction letter dated April 2, 2015, to all affected customers The letter identified the product, the problem, and the action to be taken by the customer. Winco instructed customers that they will replace the recliners at no charge to the customer. Winco will provide detailed instructions for the replacement of the seats by their facility staff. Winco will extend the remaining warranty on the entire chair for a period of 6 months as compensation. If product was further distributed customers were instructed to notify those customers of the field correction. Customers were asked to complete and return the enclosed field correction response form as soon as possible. Customers with questions should contact Customer Care Monday-Friday 8:30 AM-5:00 PM EST at 800-237-3377 or email QA-RA@wincomfg.com. For questions regarding this recall call 352-854-2929.