RecallDepth

Femoral Sponge 10 inches REF 206-715, Rx Sterile, Stryker Instruments Kalamazoo, MI. The femoral sponge is an absorbent that is used to dry the intramedullary canal prior to the application of bone cement. The suction effectively absorbs blood and fluid from the canal. It may provide swabbing/scrubbing action during suction.

This recall has been terminated (originally issued March 15, 2010).

Company
Stryker Instruments Div. Of Stryker Corporation
Recall Initiated
January 25, 2010
Posted
March 15, 2010
Terminated
October 25, 2010
Recall Number
Z-1104-2010
Quantity
9,257 all products
Firm Location
Kalamazoo, MI
Official Source
View on FDA website ↗

Reason for Recall

The firm did not have packaging validation to support the labeled shelf life. There is potential for a breach in sterility specific to the sterile packaged femoral sponge which may result in a non-sterile product.

Distribution

Worldwide Distribution -- United States, Canada, Mexico, Australia, Brazil, Switzerland, Japan, Netherlands, England, UK, Sweden, Hong Kong, Korea, Malaysia, Singapore, Argentina, Spain, and Italy.

Lot / Code Info

743288, 743289, 743330, 743331, 743332, 743333, 743362, 743363, 743364, 743365, 743378, 743379, 743380, 743381, 743390, 743391, 743400, 743402, 743403, 743404, 743405, 743418, 743419, 743420, 743421, 743422, 743423, 743445, 743446, 743447, 743448 and 743449.

Root Cause

Other

Action Taken

A "Medical Device Recall Notification" dated January 21, 2010 was issued to customers. The notification described the affected product, issue and actions to take. Consignees were asked to forward the notification to all their affected locations. Customers were instructed to immediately check inventory and quarantine any recalled product found. Consignees were instructed to destroy all product and report how much product was destroyed. For questions regarding this recall, please contact Stryker Instruments by calling 1-800-800-4236, ext. 4354

More recalls by Stryker Instruments Div. Of Stryker Corporation

View all recalls by this company →