RecallDepth

Stryker XIA 3 Polyaxial Screw, 8.5mm x 65 mm. Catalog Number: 482318565. Not Sterile; Manufactured by Stryker Spine, SA, Cestas, France. Distributed in the USA by Stryker Spine, Allendale, NJ Intended for use in the noncervical spine.

Company
Stryker Spine
Recall Initiated
December 30, 2009
Posted
February 24, 2010
Terminated
September 6, 2012
Recall Number
Z-0823-2010
Quantity
21
Firm Location
Allendale, NJ

Reason for Recall

The Xia 3 8.5mm Polyaxial screws, lot A91109 were anodized turquoise instead of Fuchsia. The turquoise color is reserved for the 9.5 mm Xia 3 screws.

Distribution

Nationwide Distribution -- IL, MO, DA, MA and CO.

Lot / Code Info

Lot number: A91109

Root Cause

Process control

Action Taken

Stryker Spine branches that received the recalled product were notified of the recall by an Urgent Product Recall letter, dated January 15, 2010, sent via Federal Express. The letter identified the affected product and the issue involved. It also discussed the potential hazards, risk mitigation, and the return process. Customers are to examine their inventory and hospital locations to identify the product. Customers are also to reconcile the product by utilizing the Customer Response Form and distribution sheet. In addition, they are to retrieve and return the affected product. Questions should be directed to Tiffani Rogers at 201-760-8206.

More recalls by Stryker Spine