Maquet Servo Ventilator SV900C/D/E (SV900) Siemens-Elema AB, Solna, Sweden

This recall has been terminated (originally issued January 5, 2010).

Company
Maquet Inc.
Recall Initiated
November 17, 2009
Posted
January 5, 2010
Terminated
June 4, 2012
Recall Number
Z-0589-2010
Quantity
29364 (8438 of those in US)
Firm Location
Wayne, NJ
Official Source
View on FDA website ↗

Reason for Recall

Some Servo Ventilators 300/300A and Servo Ventilator 900C/D/E should not be used with a system that may generate negative pressure below -100 cm H2O (Closed system suctioning) due to the risk of damaging the ventilator's pressure transducers.

Distribution

Worldwide distribution

Lot / Code Info

Serial Numbers 010001-188499 Ventilators within the specified serial number range, that are marked with one of the following labels on the inside of the lid of the pneumatic unit are exempted since they have already been fitted with a new type of the pressure transducer. 1. Upgrade EMO 12/02/1; 2. Upgrade SV300 6682-745  SV900 66 82 747.

Root Cause

Device Design

Action Taken

Maquet sent Device Correction letters commencing 11/17/09 and ending 11/23/09 by Federal Express.